HN1NVSTB | RH Insert | 1% | NIST | 5/10VDC | Temp | 100R Pt | Veris
Stainless Steel Probe +/-1% NIST 0 to 5/0 to 10 VDC 100 Ohm Platinum RTD Specialty Insertion Humidity and Temperature Sensor
Documents and manuals
Veris HN1NVSTB Documents
Overview
Details for HN1NVSTB
RH Insert | 1% | NIST | 5/10VDC | Temp | 100R Pt
HN series probe type humidity transmitters are easy to install and exceptionally accurate. Their long-term stability and trouble-free serviceability make them among the best in the industry. The electronics are embedded inside the probe, protecting them from condensation-related failures. The thin-film capacitive HS sensor elements are factory calibrated using NIST traceable calibration equipment, eliminating the need for field calibration. Field replacement of the sensor element is a snap with the patented removable sensor, lowering costs and reducing downtime.
- Sensor element - thin-film capacitive sensor element recovers from 100% saturation
- Corrosion resistant - electronics are encapsulated in stainless steel probe to resist corrosion
- Interchangeable - fully interchangeable element to 1% accuracy no calibration
- Flexible - insertion versions for application flexibility
- Compatibility - polarity insensitive 3-wire 0 to 5/0 to 10 VDC versions flexible systems compatibility
- Calibration-free interchangeable NIST traceable HS element
- RH insertion enclosure
- 1% NIST
- 3-wire, observe polarity
- Digitally profiled thin-film capacitive (32-bit mathematics) US Patent 5,844,138 HS element, this sensing element has a 1 year warranty
Specifications
Item Number | HN1NVSTB |
Brand | Veris |
Application Zone | Plant Room |
Application | Insertion Probe |
Brand | Veris |
Color | Stainless Steel |
Dimension Depth | 2" (51 mm) |
Dimension Height | 3.5" (89 mm) |
Dimension Probe Diameter | 0.5" (13 mm) |
Dimension Probe Length | 8" (203 mm) |
Dimension Width | 3.5" (89 mm) |
Electrical Wiring Size | 22 AWG (0.32 mm²) |
Feature Level | Deluxe |
Humidity Calibration Certificate | NIST |
Humidity Measurement Accuracy | ±1% |
Humidity Measurement Drift | ±1% @ 68 °F (20 °C) annually for 2 years |
Humidity Measurement Range | 0 - 100% RH |
Humidity Measurement Resolution | 0.1% RH |
Humidity Sensor Type | Thin film capacitive; replaceable |
Measurement Type | Humidity | Temperature |
Mounting | Insertion Mount |
Operating Humidity | 0-100% RH; non-condensing (will recover from saturation) |
Operating Temperature | -40 - 122 °F (-40 - 50 °C) |
Output Specifications | 0 - 5 V | 0 - 10 V |
Outputs | Analog output | Resistive temperature output |
Product Certification | CE |
Product Type | Sensor |
Range | HN |
Sensor Configuration | Insertion |
Sensor | Temp and Humidity 1% |
Supply Voltage AC Nominal | 24 VAC |
Supply Voltage DC Range | 12 - 30 VDC |
Temperature Calibration Certificate | None |
Temperature Measurement Accuracy | ±0.6 °F (±0.3 °C) |
Temperature Measurement Range | 32 - 122 °F (0 - 50 °C) |
Temperature Sensor Type | 100 Ohm Platinum RTD |
Warranty | 5 years (1 year on replaceable RH element) |