DLP-A13    | Pressure/temperature data logger | range 0-300 psia  |   Dwyer
Get Special Pricing!
Log In or Create an Account for the LOWEST PRICING on our most Popular Brands!
Brand: DwyerSKU: DLP-A13

Web Price:

$1,557.95
Availability: This product is currently available to purchase
Contact us for price breaks on large quantities!
Orders over $999 ship FREE*
The product is not cancelable or returnable unless found defective
*Free shipping applies within the continental US and weighing less than 50 lbs.
Shipping and Returns Terms
Image for Technical Expert

Speak With a Technical Expert

Call (847) 773-0645

Hours: 9am - 5pm CDT M-F

DLP-A13 | Pressure/temperature data logger | range 0-300 psia | Dwyer

Pressure/temperature data logger, range 0-300 psia

Dwyer | DLP-A13
The photo is for representation purposes only and may not be a photo of product being purchased.
Search products similar to DLP-A13

Documents and manuals

Overview

Details for DLP-A13

See catalog product page for full listing

Pressure/temperature data logger | range 0-300 psia

Series DLP Pressure/Temperature Data Logger

1/4" NPT Fitting, Up to 5000 psia

The Series DLP Data Logger can record pressure and temperature. The 1/4" NPT fitting comes standard and allows the logger to be adapted to almost any pressure fitting. The internal temperature sensor provides accurate temperature measurements without the need of a separate temperature recorder, and many of the models provide a choice between measuring pressure in psia orpsig. The DLP can easily be started and stopped from a PC or delayed to start up to six months in advance. The battery-powered data logger can store over 16,000 measurements per channel, and the easy to use DL700 software makes retrieving data simple.*DL700 interface cable required but sold separately.

Product Applications

  • Building automation
  • Clean room
  • Operating rooms
  • Process applications
  • Labs
  • Regulated environments
  • Efficient data capture tool to review processes are running correctly or to analyze alarm and out-of-spec conditions
  • Battery-powered reduces need to hard wire power allowing device to be used in a as a tool in a variety of in-and-out testing
  • PC start and stopping allows centralized control of data collection

Specifications

Series
DLP Pressure/Temperature Data Logger
Category/Type
Test Equipment
Type
Data Loggers
Series
DLP Pressure/Temperature Data Logger
Range
Temperature: -40 to 176°F (-40 to 80°C); Pressure: 0 to 30 psia(g); 0 to 100 psia(g); 0 to 300 psia(g); 0 to 500 psia(g); 0 to 1000 psia; and 0 to 5000 psia depending on the model.
Memory Size
16;383 readings per channel.
Accuracy
Temperature: ±0.9°F (±0.5°C); Pressure: 2% FSR; 0.25% at 77°F(25°C) typical.
Resolution
Temperature: 0.2°F (0.1°C); Pressure: 0.002 psia(g); 0.005 psia(g); 0.05 psia(g); 0.05 psia; and 0.2 psia depending on the model.
Sampling Method
Stop on memory full.
Sampling Rate
Selectable from 2 sec to 12 hrs.
Computer Requirements
Windows® 95; Windows® 98; Windows® 2000; Windows® ME; Windows NT®; or Windows® XP operating system; and one free USB port.
Power Requirements
3.6 V TL2150 lithium metal battery; installed functional; user replaceable.
Battery Life
1 yr (approx).
Interface
USB port (interface cable required).
Material
Stainless steel.
Weight
12 oz (340 g).
Agency Approvals
CE.

Reviews

People Also Searched
DLP-A13
Dwyer DLP-A13
DLP A13
DLPA13
Dwyer DLP A13
Dwyer DLPA13
DLP-A13 price
Dwyer DLP-A13 price
DLP A13 price
DLPA13 price
Dwyer DLP A13 price
Dwyer DLPA13 price