CAM-20 | Compressed air meter | 1/2
Get your Special Pricing!
Log In / Create an Account for the LOWEST PRICING on our most Popular Brands!

Web Price:

$1,988.27
Availability: This product is currently available to purchase

Contact us for price breaks on large quantities!

Orders over $999 ship FREE*
*Free shipping applies within the continental US and weighing less than 50 lbs. Shipping and Returns Terms
Ship to Estimates
Are you from Chicago?
Yes / No
This item ships to
Change
Cancel


*shipping cost are estimated until you complete your order

The product is not cancelable or returnable unless found defective

Speak With a Technical Expert

Image for Technical Expert

CAM-20 | Compressed air meter | 1/2" NPT | range 0.2-44.1 SCFM. | Dwyer

Compressed air meter, 1/2" NPT, range 0.2-44.1 SCFM.

Brand: DwyerSKU: CAM-20
Dwyer | CAM-20
The photo is for representation purposes only and may not be a photo of product being purchased.
Search products similar to CAM-20

Overview Details for CAM-20

See catalog product page for full listing

Compressed air meter | 1/2" NPT | range 0.2-44.1 SCFM.

Series CAM Compressed Air Meter

Steady-State and Cumulative Totals PNP/4-20 mA Output

The Series CAM Compressed Air Meter operates by the calorimetric measuring principle to detect the standard volume flow of operating compressed air. It is designed specifically to evaluate the current flow rate and the current consumed quantity in compressed air systems. With an easily programmable digital display, this unit offers a quick, accurate readout. The Series CAM is constructed of high-grade materials and is virtually maintenance free.

Specifications

Series
CAM Compressed Air Meter
Category/Type
Flow
Type
Flow Transmitters
Series
CAM Compressed Air Meter
Service
Compressed air; air quality (see chart on Catalog Page).
Wetted Materials
Stainless steel (304S15); ceramics; glass passivated; PEEK; polyester; fluoroelastomer; anodized aluminum.
Accuracy
Class 141: ±3% of measured value + 0.3% of full scale value; Class 344: ±6% of measured value + 0.6% of full scale value.
Response Time
<0.1 seconds.
Temperature Limit
32 to 140°F (0 to 60°C).
Pressure Limit
232 psig (16 bar).
Humidity Limit
90% RH.
Power Requirements
19-30 DC.
Output Signal
Output 1: PNP Open Collector: VOUT(max)=30 VDC/ISINK(max)= 250 mA; Output 2: 4-20 mA (scaleable) or PNP Open Collector.
Loop Resistance
500 ohms.
Current Consumption
<100 mA.
Electrical Connections
4-pin M12 (micro) cable connection.
Process Connections
1/2"; 1" and 2" NPT.
Display
4-digit alpha numeric red LED.
Enclosure Rating
IP 65 (NEMA 4X).
Mounting Orientation
Vertical; horizontal; side with pipe length facing left.
Weight
CAM-10: 12.5 lb; CAM-20: 2.35 lb; CAM-30: 4.16 lb.
Agency Approvals
CE.

Reviews