Scroll to view all
Get your Special Pricing!
Documents: 477AV-3
477AV-3 | Digital manometer | range 0-200.00" w.c. | air velocity/flow modes. | Dwyer
Digital manometer, range 0-200.00" w.c., air velocity/flow modes.
The photo is for representation purposes only and may not be a photo of product being purchased.
Search products similar to 477AV-3
Overview Details for 477AV-3
See catalog product page for full listing
Digital manometer | range 0-200.00" w.c. | air velocity/flow modes.
Series 477AV Handheld Digital Manometer
Measures Pressure, Flow and Velocity; ±0.5% Accuracy
The Series 477AV Handheld Digital Manometer is now available with pressure, flow, and velocity measurements along with a number of other convenient features. The built-in air velocity and flow calculations provide accuracy and conserve time and error associated with manual calculations. Also featured on this unit are adjustable zero and span values for calibrating in the field, as well as a damping feature to compensate for the fluctuating of readings.
NIST CALIBRATION available in Product Configurator.
Product Applications
- Air flow monitoring, when used with a Dwyer® pitot tube
- Duct static pressure
- Commercial air balancing
- Building-zone pressure
- Calculates and displays air velocity and volumetric air flow
- Rugged aluminum case protects instrument from damage during transport/use
- 9 selectable English and metric engineering units
- Large, easy to read display with backlight for use in dark areas
- Stores up to 40 readings with minimum, maximum and average statistics
- Convenient all-in-one air velocity kit option available
Specifications
- Series
- 477AV Handheld Digital Manometer
- Category/Type
- Pressure
- Type
- Manometers
- Series
- 477AV Handheld Digital Manometer
- Service
- Air and compatible gases.
- Wetted Materials
- Consult factory.
- Accuracy
- ±0.5% FS; 60 to 78°F (15.6to 25.6°C); ±1.5% FS from 32 to 60°Fand 78 to 104°F (0 to 15.6°C and 25.6to 40.0°C).
- Pressure Hysteresis
- ±0.1% of FS.
- Pressure Limits
- See chart on catalog page.
- Temperature Limits
- 0 to 140°F (-17.8 to 60°C).
- Compensated Temperature Limits
- 32 to 104°F (0 to 40°C).
- Storage Temperature Limits
- -4 to 176°F (-20 to 80°C).
- Display
- 0.42" (10.6 mm) 4 digit LCD.
- Resolution
- See chart on catalog page.
- Units of Pressure
- in w.c.; ft. w.c.; inHg; psi; oz/in²; mm w.c.; cm w.c.; mm Hg;mbar; Pa; kPa; hPa.
- Units of Velocity
- fpm; fps; mph; m/h; m/s; k/h; knot.
- Units of Flow
- cfm; m3/h; m3/s.
- Power Requirements
- 9 V Alkaline battery; installed non-functional; user replaceable.
- Weight
- 10.2 oz (289 g).
- Process Connections
- Two barbed connections for use with 1/8" (3.18mm) or 3/16" (4.76 mm) ID tubing. Two compression fittings for use with 1/8"(3.18 mm) ID x 1/4" (6.35 mm) OD tubing for 477AV-7 and 477AV-8 only.
- Compliance
- CE.